Appendix A
standard deviation. Plus or minus one standard deviation encompasses 68%
of the population.
‡The Masimo SET
®
SpO
2
parameter has been validated for low perfusion
accuracy in bench top testing against a Biotek Index 2 simulator and
Masimo’s simulator with signal strengths of greater than 0.02% and a %
transmission of greater than 5% for saturations ranging from 70 to 100%.
This variation equals plus or minus, one standard deviation. Plus or minus
one standard deviation encompasses 68% of the population.
Masimo
®
Sensor Accuracy
Sensor Model
SpO
2
Range 70% - 100%
LNOP
LNOP-ADT ±2 digits
LNOP-ADT Long ±2 digits
LNOP-PDT ±2 digits
LNOP-NEO ±3 digits
LNOP-NEO PT ±3 digits
LNOP-DCI (reusable) ±2 digits
LNOP-DCSC (reusable) ±2 digits
LNOP-DCIP (reusable) ±2 digits
NRI25 (reusable) ±2 digits
Resolution
Saturation (% SpO
2
)1%
Pulse Rate (bpm) 1
Low Perfusion Performance
>0.02% Pulse Amplitude Saturation (% SpO
2
) ±2 digits
and % Transmission >5% Pulse Rate ±3 digits
Interfering Substances
Carboxyhemoglobin may erroneously increase readings. The
level of increase is approximately equal to the amount of
carboxyhemoglobin present. Dyes, or any substance
containing dyes, that change usual arterial pigmentation may
cause erroneous readings.
Sensor Light Source
Wavenlength Infrared: 905 nm (nominal)
Red: 660 nm (nominal)
Power Dissipation Infrared: 22.5 mW (max)
Red: 27.5 mW (max)