Using the Monitor
When a monitor mode measurement is in progress, it is
terminated when the probe is inserted into the probe holster.
Note: These temperature readings are not stored in trends
and not reported via host comms.
General Warning
The performance of the Monitor may be degraded if it
is operated outside of the environmental conditions
specified in Appendix A.
General Cautions
Be careful not to overextend the coiled cord of the
temperature probe. Overextension can damage the
probe coil connector interfaces.
Accurate oral temperatures (blue) can only be obtained
by placing the probe under the tongue in the right or
left sublingual pocket. Temperatures in other locations
in the mouth can vary by more than 2 °F or 1 °C.
Accurate rectal temperatures can only be obtained by
using the red temperature probe. Red and blue
temperature probes are not interchangeable.
Do not allow the tip of the predictive temperature
probe to come into contact with a heat source (e.g.,
hands or fingers) prior to taking a temperature
determination. If this occurs, allow 5 seconds for the
probe tip to cool before proceeding.
Electromagnetic Compatibility: Operating the
thermometer near equipment which radiates high-
energy electromagnetic and radio frequencies
(electrosurgical/cauterizing equipment, portable
radios, cellular telephones, etc.) may cause false alarm
conditions. If this happens, reposition the thermometer
away from the source of interference and perform a
new measurement.
Use only IVAC
®
Probes and P850A probe covers with
the TURBO
TEMP™ Thermometer. The size, shape,
and thermal characteristics of the probe covers can
affect the performance of the instrument. Inaccurate