72
Clinician Menu
Press
. Selection of this button displays a panel for setting the
default BP target inflation pressure. Adjusting the default
target pressure will automatically update the current inflation
target pressure and will be used for the next reading. The
range of adjustment is 100 mmHg to 180 mmHg, and the
setting is retained when the Monitor is turned off.
The initial target pressure can be set between 100 and 180
mmHg in 5 mmHg increments. The factory default is 150
mmHg for adults and 110 for neonates. This is indicated by
the “AUTO” label at the end of the adjustable range. When
the target pressure is changed, the next determination will
use the new target inflation value if no systolic is available.
When adjusted under the Clinician menu of the Service
mode, the target pressure is adjusted permanently.
Retain Values. The Retain Values setting indicates the
amount of time that the manual BP and Temp measurement
values stay on screen. This is also the setting for the amount
of time that Spot Mode values are displayed after a freeze
period.
OK. Selection of this button returns the user to the Clinician
menu.
Info
. Selection of this button causes the most recent
calibration date of the NIBP system to be displayed.
Selection of OK returns the user to the Clinician menu.