normal SpO
2
reporting of values when several consecutive
good pulse determinations are made.
General Warnings
Do not use the SpO
2
function during magnetic
resonance imaging (MRI). Adverse reactions include
potential burns to patients as a result of contact with
attachments heated by the MRI radio frequency pulse,
potential degradation of the magnetic resonance
image, and potential reduced accuracy of SpO
2
measurements. Always remove oximetry devices and
attachments from the MRI environment before
scanning a patient.
The use of cardio-green and other intravascular dyes at
certain concentrations may affect the accuracy of the
SpO
2
measurement.
•The SpO
2
function is calibrated to read functional
arterial oxygen saturation. Significant levels of
dysfunctional hemoglobins such as carboxyhemoglobin
or methemoglobin may affect the accuracy of the SpO
2
measurement.
Pulse oximetry readings and pulse signals can be
affected by certain environmental conditions, sensor
application errors, and certain patient conditions. See
the appropriate sections of this manual for specific
safety information.
The use of accessories, transducers, and cables other
than those specified may result in increased emission
and/or decreased immunity and inaccurate readings of
the Monitor.
General Cautions
As with any clip-on sensor, pressure is exerted. The
clinician should be cautious in using a clip-on sensor on
patients with compromised circulation (e.g., because of
peripheral vascular disease or vasoconstricting
medications).