4-14 PRO 100-400 V2 Patient Monitor Service Manual Revision A
2009829-001
Calibration & Maintenance: SECTION 4. GENERAL MAINTENANCE
Check Button
This button allows the NIBP calibration to be checked. In this mode,
the Monitor functions as a digital manometer. The systolic LED
displays the output of PT1, and the diastolic LED displays the output
of PT2.
With no hose attached, the systolic and diastolic displays will indicate
“000.” See Section 4.7 for a more detailed description of this mode
and the calibration procedure.
Cal Button
This button initiates the calibration procedure. See Section 4.7 for a
step-by-step description of this procedure.
Pop Off Button
This button functions in a similar manner to the Check button, but the
overpressure point is set to the neonate value. See Section 4.7 for a
more detailed description of using this function.
OK Button
This button returns the display to the Main Service menu.
4.5.4. System Menu
The System menu, pictured above, accesses the options that modify
the basic configuration of the PRO Monitor, test the EEPROM, and
print the error log.
Language Button
This button displays the language choices. The language dialog
boxes are not pictured. The PRO Monitor software can display menus
in five languages: UK, USA, German, French, and Spanish.
Selecting any language button removes all other language buttons
from the screen, indicating that the remaining button is the chosen
language. Selecting Clear restores all the language buttons, allowing
the user to select again. OK saves the selection and a dialog box
requests that the Monitor be turned off. If no language is selected
when OK is pressed, the language is stored as undefined. In this
Przeglądanie stron 65