2-8 PRO 100-400 V2 Patient Monitor Service Manual Revision A
2009829-001
Product Description: SECTION 2. PRODUCT DESCRIPTION
16. ON/OFF switch: Controls on/off state of PRO Monitor; push for
power on and push again for power off.
17. Battery power indicator: Yellow LED indicates operation and
charge status of battery.
18. SpO
2
sensor connector: SpO
2
sensor extension cable attaches
here (Models 300 and 400).
19. Mean arterial pressure display: 3-digit red LED indicates
measured MAP in mmHg and shows instantaneous cuff pressure
during BP determination.
20. SpO
2
pulse indicator: Yellow LED in heart symbol flashes to
indicate that real-time pulse rate measurements are being derived
from SpO
2
signals (Models 300 and 400).
21. Rotor: Used to highlight and select items in LCD menus; if monitor
is off, pressing rotor will switch monitor on.
22. Pulse BPM display: 3-digit yellow LED shows pulse rate in beats
per minute.
23. SpO
2
display: 3-digit red LED indicates oxygen saturation in %
(Models 300 and 400).
24.SpO
2
motion/artifact indicator LED: For NELLCOR, LED
Illuminates when motion artifact is detected (Models 300 and 400).
For Masimo, LED illuminates when low perfusion or low signal quality
is detected (Models 300 and 400).
25. LCD (liquid crystal display): Displays all alarms, user interface
messages, and configuration options.
26. Alarm silence switch: Alternately mutes and enables audible
alarms; when pushed once after alarm sounds (silence on), the
switch illuminates to indicate that audible alarms have been silenced
for 2 minutes.
27. AUTO BP key: Press to start Auto BP mode.
28. Light sensor: Automatically measures ambient light to set LED
display intensity.
29. START/STOP BP key: Press to start or stop a BP, Auto, Stat, or
Vitals determination.
30. Cuff connector: BP cuff hose attaches here.
Not Shown: The printer access door is located on the left side of the
monitor. Push the latch to open the door and access the paper supply
and print head. For details, see Section 5, Component Information.